Troubleshooting Guide
221
If the amplitude is not identical with the actual voltage
Check that the attenuation factor of the probe is identical
with the attenuation factor set in the channel menu.
If instrument is not connected over LAN
Try using DHCP if not already.
Reboot the instrument, then try again.
How do I extract the deep memory from the DSO?
This can only be extracted using provided PC software.
It cannot be exported directly to a connected front USB
flash drive.